Skip to main content
Log in

A Newly Developed Video-Meibography System Featuring a Newly Designed Probe

  • BRIEF COMMUNICATION
  • Published:
Japanese Journal of Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To demonstrate the usefulness of a video-meibography system featuring our newly designed probe.

Methods

This study utilized a newly designed probe featuring an obliquely inclined, T-shaped head containing 16 windows in the portion of the probe applied for emitting infrared light. The transilluminated structure of the meibomian gland is conducted to a highly sensitive, infrared charge-coupled device video camera and recorded digitally. The utility of the new probe was compared with the intraocular illumination probe traditionally used for meibography, both being applied to the eyelids of ten healthy volunteers and ten patients with meibomian gland dysfunction. The results were then analyzed.

Results

The newly designed probe facilitated eyelid eversion and transilluminated the meibomian gland structure effectively. Moreover, subjects experienced no irritation and only limited invasiveness when the new probe was applied.

Conclusions

Our meibography system featuring the newly designed probe provided greater ease of use, an improved visualization of the meibomian gland structure, and increased comfort for patients during the examination. Jpn J Ophthalmol 2007;51:53–56 © Japanese Ophthalmological Society 2007

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

References

  1. R Tapie (1977) ArticleTitleEtude biomicroscopique des glandes de meibomius Ann Oculistique 210 637–648

    Google Scholar 

  2. JB Robin JV Jester J Nobe N Nicolaides RE Smith (1985) ArticleTitleIn vivo transillumination biomicroscopy and photography of meibomian gland dysfunction. A clinical study Ophthalmology 92 1423–1426 Occurrence Handle4069605 Occurrence Handle1:STN:280:BimD28nlt1Y%3D

    PubMed  CAS  Google Scholar 

  3. WD Mathers WJ Shields MS Sachdev WM Petroll JV Jester (1991) ArticleTitleMeibomian gland dysfunction in chronic blepharitis Cornea 10 277–285 Occurrence Handle1889213 Occurrence Handle10.1097/00003226-199107000-00001 Occurrence Handle1:STN:280:By6A28vkt1I%3D

    Article  PubMed  CAS  Google Scholar 

  4. WD Mathers T Daley R Verdick (1994) ArticleTitleVideo imaging of the meibomian gland Arch Ophthalmol 112 448–449 Occurrence Handle8155046 Occurrence Handle1:STN:280:ByuB3cnivVU%3D

    PubMed  CAS  Google Scholar 

  5. N Yokoi F Mossa JM Tiffany AJ Bron (1999) ArticleTitleAssessment of meibomian gland function in dry eye using meibometry Arch Ophthalmol 117 723–729 Occurrence Handle10369581 Occurrence Handle1:STN:280:DyaK1M3ps1aksA%3D%3D

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Norihiko Yokoi.

About this article

Cite this article

Yokoi, N., Komuro, A., Yamada, H. et al. A Newly Developed Video-Meibography System Featuring a Newly Designed Probe. Jpn J Ophthalmol 51, 53–56 (2007). https://doi.org/10.1007/s10384-006-0397-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10384-006-0397-y

Key words