Introduction
Diabetic retinopathy (DR) is the most common microvascular complication of diabetes mellitus. It is a major cause of vision impairment and blindness.1 Retinal screening and referral for treatment for those identified having DR can prevent vision loss.2–5 For this reason, many countries are introducing DR screening and treatment programmes.6–8
A recent systematic review of DR screening found that in low-income and middle-income countries (LMIC), common barriers include limited skilled human resources and lack of access to eye facilities.9 Use of artificial intelligence (AI) for grading of retinal images could help to reduce the time spent by ophthalmic specialists reviewing images.10 11 AI in DR screening can allow quick assessment of a large number of images and communication of the decision to refer, or not, to the patients at the point of care, and in the last few years these technologies have started to be validated.12–14 As the quality of smartphone cameras improves, there has been investment and research into their use as portable retinal cameras, offering a lower cost and transportable option in low resource and rural settings.15
Four recent meta-analyses reported sensitivities for AI to grade DR between 87% and 97%.16–19 Most studies reported AI systems which used datasets from high-quality images taken with state-of-the-art retinal cameras in eye clinic settings. Some studies, including a large-scale real-world use of AI in Thailand, have assessed community screening in field settings, reporting sensitivities between 84% and 91% for referable DR and 91% for vision threatening DR.20–22
The prevalence of diabetes in the adult population in Dominica is estimated to be 17.7%.23 Dominica has been screening for DR since 2005, but its programme coverage is limited with approximately 1500 of the estimated 7000 adults living with diabetes being screened each year. There are two employed ophthalmic technicians in the public sector in Dominica certified to grade retinal images, but their time to screen DR is limited by other clinical duties. There are two retinal cameras, one fixed (Centervue DRS) in a hospital in Roseau, the capital, and a smartphone camera (Remidio) used in a mobile clinic that visits rural districts. The ophthalmology services in Dominica are equipped to deliver treatment to patients with vision threatening DR.
AI-assisted grading in the mobile clinic could help overcome human resources constraints and increase DR screening coverage. There is an AI software application that can be used offline with the smartphone-based ‘Fundus on Phone’ retinal camera currently used in Dominica.24 Studies in India using this AI software and camera have reported a sensitivity of 83% to detect any DR, and a sensitivity of 93% to detect ‘referable’ DR.25–27
This study aimed to evaluate the diagnostic accuracy of Medios AI software for the diagnosis of referable diabetic retinopathy (RDR) using mydriatic retinal images when deployed and integrated in a real-world DR screening programme in a Caribbean population in Dominica.